EasyManua.ls Logo

Siemens Rapidlab 1200 - Page 422

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-62 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
3 Syringe 42 0.012 1.33 1.36 97.7 0.9
4 Syringe 45 0.022 1.91 1.93 99.0 1.1
5 Syringe 42 0.043 2.30 2.36 97.8 1.9
6 Syringe 45 0.062 4.24 4.09 103.7 1.5
1Micro-
capillary
11 0.015 0.57 0.56 100.5 2.6
2Micro-
capillary
12 0.016 1.24 1.24 100.0 1.3
3Micro-
capillary
12 0.005 1.29 1.30 99.8 0.4
4Micro-
capillary
12 0.044 1.97 1.98 99.5 2.2
5Micro-
capillary
12 0.046 2.36 2.38 99.4 1.9
6Micro-
capillary
12 0.071 3.97 3.98 99.8 1.8
1Micro-
syringe
25 0.023 0.51 0.56 91.1 4.5
2Micro-
syringe
24 0.012 1.26 1.24 101.4 1.0
3Micro-
syringe
24 0.016 1.30 1.30 100.4 1.3
4Micro-
syringe
24 0.026 2.00 1.98 101.1 1.3
5Micro-
syringe
24 0.044 2.43 2.38 102.2 1.8
6Micro-
syringe
24 0.121 4.20 3.98 105.7 2.9
* WRSD = within-run standard deviation
Table E-61 Rapidlab 1265 Recovery and Precision Testing—Glucose
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 21 0.87 48 45 106.9 1.8
2 Capillary 21 2.55 90 89 101.0 2.8
3 Capillary 21 2.04 210 210 100.0 1.0
4 Capillary 21 3.56 326 349 93.4 1.1
5 Capillary 21 3.66 450 470 95.8 0.8
1 Syringe 39 1.37 42 38 110.8 3.3
Table E-60 Rapidlab 1265 Recovery and Precision Testing—Ca
++
Level Mode N WRSD
*
Observed Expected Recovery WRCV

Table of Contents

Related product manuals