Rapidlab 1200 Operator’s Guide: System Specifications E-63
02087462 Rev.V
2 Syringe 40 1.57 87 82 105.5 1.8
3 Syringe 45 2.02 214 211 101.2 0.9
4 Syringe 45 2.24 339 358 94.9 0.7
5 Syringe 42 5.25 477 472 101.1 1.1
1Micro-
capillary
12 2.02 43 44 96.8 4.7
2Micro-
capillary
12 2.47 86 88 97.9 2.9
3Micro-
capillary
12 1.41 224 219 102.4 0.6
4Micro-
capillary
12 7.42 332 343 96.7 2.2
5Micro-
capillary
12 4.21 447 452 98.8 0.9
1Micro-
syringe
24 1.76 48 48 101.0 3.7
2Micro-
syringe
24 2.05 99 98 101.1 2.1
3Micro-
syringe
24 1.87 226 222 102.1 0.8
4Micro-
syringe
24 18.80 367 374 98.1 5.1
5Micro-
syringe
24 8.40 456 452 100.9 1.8
1 pH, Glu,
Lac
Capillary
21 0.83 52 44 116.5 1.6
2 pH, Glu,
Lac
Capillary
24 1.80 98 93 106.3 1.8
3 pH, Glu,
Lac
Capillary
24 2.60 203 205 99.4 1.3
4 pH, Glu,
Lac
Capillary
24 3.93 355 365 97.3 1.1
5 pH, Glu,
Lac
Capillary
21 9.84 495 526 94.2 2.0
Table E-61 Rapidlab 1265 Recovery and Precision Testing—Glucose
Level Mode N WRSD
*
Observed Expected Recovery WRCV