E-64 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
1 pH, Glu,
Lac
Syringe
48 1.13 51 43 118.9 2.2
2 pH, Glu,
Lac
Syringe
48 2.57 96 87 110.0 2.7
3 pH, Glu,
Lac
Syringe
45 4.67 215 215 99.7 2.2
4 pH, Glu,
Lac
Syringe
46 5.21 371 384 96.4 1.4
5 pH, Glu,
Lac
Syringe
48 5.71 494 520 95.0 1.2
* WRSD = within-run standard deviation
Table E-62 Rapidlab 1265 Recovery and Precision Testing—Lactate
Level Mode N WRSD
*
Observed Expected Recovery WRCV
1 Capillary 18 0.06 1.6 1.6 102.9 3.5
2 Capillary 21 0.04 2.1 2.2 95.4 2.1
3 Capillary 18 0.10 5.1 5.2 96.3 1.9
4 Capillary 21 0.12 9.6 11.0 87.4 1.2
5 Capillary 18 0.17 12.7 14.4 88.1 1.3
1 Syringe 39 0.04 1.2 1.3 96.2 3.5
2 Syringe 42 0.05 2.3 2.2 101.5 2.1
3 Syringe 38 0.10 5.5 5.6 98.4 1.8
4 Syringe 45 0.15 9.9 10.9 90.8 1.5
5 Syringe 36 0.26 13.4 14.4 92.9 1.9
1Micro-
capillary
12 0.05 1.1 1.0 104.4 4.7
2Micro-
capillary
12 0.12 4.1 4.4 93.4 2.9
3Micro-
capillary
12 0.07 5.8 6.7 87.1 1.2
4Micro-
capillary
12 0.06 9.5 11.5 82.5 0.7
5Micro-
capillary
12 0.51 13.1 16.2 80.9 3.9
Table E-61 Rapidlab 1265 Recovery and Precision Testing—Glucose
Level Mode N WRSD
*
Observed Expected Recovery WRCV