EasyManua.ls Logo

Siemens Rapidlab 1200 - Page 425

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rapidlab 1200 Operator’s Guide: System Specifications E-65
02087462 Rev.V
1Micro-
syringe
24 0.06 1.2 1.2 98.6 5.1
2Micro-
syringe
24 0.06 3.3 3.6 90.6 2.0
3Micro-
syringe
24 0.14 5.9 6.4 93.0 2.3
4Micro-
syringe
24 0.23 9.7 10.8 89.4 2.3
5Micro-
syringe
24 0.23 11.9 13.9 85.9 2.0
1 pH, Glu,
Lac
Capillary
21 0.05 0.9 0.9 109.7 5.1
2 pH, Glu,
Lac
Capillary
21 0.07 2.8 2.8 98.5 2.6
3 pH, Glu,
Lac
Capillary
15 0.06 4.9 5.3 92.9 1.3
4 pH, Glu,
Lac
Capillary
21 0.13 10.5 12.1 86.4 1.2
5 pH, Glu,
Lac
Capillary
15 0.25 14.2 17.4 81.5 1.7
1 pH, Glu,
Lac
Syringe
41 0.03 0.9 0.8 106.5 3.1
2 pH, Glu,
Lac
Syringe
45 0.08 2.5 2.7 93.5 3.2
3 pH, Glu,
Lac
Syringe
45 0.11 5.5 6.0 91.4 1.9
4 pH, Glu,
Lac
Syringe
42 0.33 10.8 12.5 86.9 3.1
5 pH, Glu,
Lac
Syringe
45 0.34 14.8 17.9 82.8 2.3
* WRSD = within-run standard deviation
Table E-62 Rapidlab 1265 Recovery and Precision Testing—Lactate
Level Mode N WRSD
*
Observed Expected Recovery WRCV

Table of Contents