EasyManua.ls Logo

Agilent Technologies 7500 ICP-MS

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-60
Agilent 7500 ICP-MS ChemStation Operators Manual
Tuning
Scan line of Octopole
Three parameters are used to determine this scan line:
Ramp Offset: The Offset voltage of the RF voltage in the mass portion area.
Bend Mass: The Mass number moving from the mass portion area to the constant
voltage area.
OctP RF: The RF voltage in the constant mass area.
OctP RF can be set in tuning parameters in the ICP-MS Tuning Window.
To enable the scan and set the two remaining parameters you use the Octopole
Setup Dialog Box.
To set these parameters, carry out the following steps:
1 Select Instrument>>Instrument Control in the ICP-MS Top Window, and display
the ICP-MS Instrument Control Window
2 Select Maintenance>>Octopole in the Instrument Control Window.
The Octopole Setup Dialog Box will appear.
Ramp Offset
RF Voltage
Bend Mass
OctP RF
(Tuning parameter)
QP Mass

Table of Contents

Related product manuals