4-74
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Tuning
2 Adjustment of H
2
gas flow
The higher the flow of H
2
gas, the lower the intensity of the sample and the
background. Adjust the flow in order to make the background low enough and
to get a necessary level of sensitivity. To optimize the flow rate of H
2
gas use
the ICP-MS Tuning-Reaction Gas Window as follows:
1) Select Tune>>Reaction Gas in the ICP-MS Tuning Window.
The ICP-MS Tuning-Reaction Gas Window will appear.
In this window, the data of the tuning and blank solutions are collected as the gas
flow change. The Graph button will produce a graph of the sensitivity and BEC
(Background Equivalent Concentration). This graph will help to determine the
flow rate of H
2
gas.
Reaction Gas Tuning Window
2) Aspirate the tuning solution.
3) Set the mass number and integration time by selecting Acq.Params.
4) Input the concentration of each mass number of the tuning solution in Std
Conc Text Box and select the concentration unit in Conc Unit.
5) Select Std (standard solution) in Current Solution.
6) Select H
2
Gas in the Ramp Flow group and input the minimum value,
maximum value, and the step size of the H
2
gas flow rate to adjust the Flow
Rate.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals