4-77
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Tuning
NOTE Once the parameter of the reaction gas flow rate is tuned; only a fine adjustment of
the flow rate is enough for daily tuning.
He Mode (When Helium is used as the Reaction Gas)
It is recommended to tune in He mode after tuning in standard mode.
Autotune cannot be used in He mode.
The following is an example of the sensitivity adjustment method in He mode: This
assumes that adjustment in standard mode is complete. First, tune the lens voltage
with an appropriate He flow, and then adjust the He flow for optimum conditions.
1 Adjustment of lens voltage
a.Save the tuning parameters from standard mode.
b.Aspirate the tuning solution and confirm that Co (59 amu), Y (89 amu), and
Tl (205 amu) show normal counts in the ICP-MS Tuning-Sensitivity
Window.
c.Turn Reaction Mode Check Box ON and set the value 3.5 mL/min. for the
He Gas flow rate.
d.Enter the typical tune value of each parameter. But use the value from the
Standard Mode tune;
• RF Power
• Sampling Depth
• Carrier Gas
• Makeup Gas
• Torch-H, V
• Extract 1
• Extract 2
e.Maximize the sensitivity of Co (59 amu) by adjusting the Omega Lens-cs,
QP Focus, and OctP RF.
f. Adjust the Makeup Gas and choose a value that gives maximum sensitivity.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals