4-86
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Tuning
Octopole Setup Dialog Box
3 Set each parameter necessary to scan the Octopole
• Enable Scan: To scan, input a check mark in this check box. If there is no check
mark in the check box, Octopole will not be scanned. Usually, leave the check
box blank.
• Ramp Offset: Input the offset voltage of the RF voltage in the mass portion area.
The range is 30V to 200V. Input a value lower than the OctP RF in the tuning
parameter.
• Bend Mass: Input the mass number moving from the mass portion area to the
constant voltage area. The range is 1 amu to 260amu. The Mass number input in
the bend mass will be included in the constant voltage area. To make all the mass
range the mass portion area, input 260 amu.
4Click OK after finishing the parameter setting.
The setting is complete and the Octopole Setup Dialog Box closes.