4-89
Agilent 7500 ICP-MS ChemStation Operators Manual
Tuning
mass toward higher mass.
Axis Offset
Adjusts mass calibration. The higher the value, shifts the peak position of all
masses toward higher mass.
QP Bias
Controls the speed of ions as they pass through the Q-pole. A typical value is 0 V
without the ShieldTorch system. It may be used up to 6 V when the peak shape
and resolution are not good. This voltage should be the same as the Plate Bias to
avoid higher background. A typical value with the ShieldTorch installed will be
-10 to 0 V.