4-94
Agilent 7500 ICP-MS ChemStation Operators Manual
Tuning
For P/A Factor tuning, complete the following steps:
1 Aspirate tuning solution.
2 Select Tune>>P/A Factor.
The P/A Factor Tuning dialog box appears.
3 Set the mass number and/or elements to be used for tuning.
The mass number and/or element can also be loaded from the current method by
clicking Load masses from Acq. method. Having a method specifically for P/A
Factor tuning is recommended. For information about creating a method, see
Chapter 5, “Creating a Method”.
Without Autosampler With Autosampler
P/A Factor Tuning Dialog Box
4 Cancel check of the Merge in the current data box, if necessary.
The Merge in the current data box is disabled when adjusting the P/A factor for the
first time. It is usually checked for the following times. Cancel the check when the
EM adjustment has been carried out.
5 Enter the vial number of the autosampler in the ALS Tuning Solution Vial.
When the autosampler is set in the Configuration task, this box appears. Enter the
vial number of the tuning solution. When “0” is entered, the tuning begins without

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals