Agilent 7500 ICP-MS ChemStation Operators Manual
Setting Up a Sequence
3) Set an LC/GC vial number in the LC/GC Vial field.
If you leave the LC/GC Vial field blank, the method in the LC/GC ChemStation
will not be executed. You can set a vial number from -1 to 200. (The range of
numbers you can set varies depending on the instrument connected.)
12 Repeat the above steps for each sample to complete the Sample Log Table.
You can use the popup menu by clicking the right mouse to make operation easier.
Any numeric portion except the method name can be automatically incremented by
using Fill down or Repeat.
Popup Menu for Setting
NOTE The 4-digit vial number for the standard or sample racks is incremented
corresponding to the current autosampler configuration. The 1 digit vial number for
the water, blank, and tuning solution rack is incremented up to 5.