EasyManuals Logo

Agilent Technologies 7500 ICP-MS User Manual

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #457 background imageLoading...
Page #457 background image
12-21
Agilent 7500 ICP-MS ChemStation Operators Manual
Performing a Quantitative Analysis
1 Select the Graph Detail tab in the Edit Calibration dialog box.
The Graph Detail dialog box will appear.
Graph Detail Dialog Box
The left side of the dialog box shows the calibration information and a calibration
curve. Clicking [-] will zoom out and clicking [+] will zoom in. Clicking [<--] will
reset the zoom and clicking [++>] results in maximum zoom.
The table on the right side shows the concentration, calculated concentration value,
CPS or count, strength ratio (if internal standard correction is used), and detector
mode (P or A) for each level, as well as the RSD table for the element at each
calibration level.
2 To check a point of a level on the calibration curve, perform the following
procedure.
1) Select a line in the table on the right side.
The selected line will be highlighted and a circle will appear around a point on the
graph.
The color of the point on the calibration curve provides an indicator as follows: light
blue for pulse mode and green for analog mode.
2) To reject a point on the calibration curve, double-click on the Rjct field. The
Reject field will show “*.”

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals