2-8
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Configuration
9 Match the ChemStation configuration with the Instrument Hardware.
The Miscellaneous dialog box appears when the Miscellaneous button is clicked.
Miscellaneous Dialog Box
• Hardware Option
SC Cooling:
In normal operation when introducing a liquid sample, the spray chamber is
cooled to stabilize the fine particles supplied to the plasma. For other sample
introduction systems like laser ablation, the connection is made directly to the
torch and the spray chamber is not used. When cooling the spray chamber, select
SC Cooling. When not cooling the spray chamber, the selection is not required.
Pump2 (Option):
It is selected when a second (optional) peristaltic pump is installed on the
ICP-MS. When there is no option pump this box is not selected.