3-25
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Startup, Shutdown and Status
4 In Standby mode
5In Analysis mode
Parameters Limit Conditions
Analyzer pressure (AN) > 2 x 10
-3
Pa Agilent 7500 goes back to Shutdown mode.
Backing pressure (BK) > 40 Pa Agilent 7500 goes back to Shutdown mode.
Parameters Limit Conditions
Interface pressure (IF) > 530 Pa Agilent 7500 goes back to Standby mode.
Analyzer pressure (AN) > 2 x 10
-3
Pa Agilent 7500 goes back to Standby mode.
Backing pressure (BK) > 40 Pa Agilent 7500 goes back to Shutdown mode.