GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
3-66 Section 3-14 - Agent Configuration (R3 only)
Use Agent Configuration to:
Edit and configure the following:
- Enterprise server, Select “Production” for install base. Select “Staging”for test purpose
- Enterprise host name in the agent
- Enterprise port number in the agent
- Proxy server, Proxy port in the agent, Consult institutions’ IT personnel for Proxy settings
- CRM number in the agent. The default CRM Number is same as the serial number prefixed with
"LP9P7".
-Display name in the agent
Enter the user name and password for the proxy
Reset the edited unsaved value
Update contact details
For more information, see "Agent Configuration" of Common Service Desktop (R3 only) section on
Page 3-74.
NOTICE
!! NOTICE:
If you want to activate Request for Service (RFS), you need the verified CRM Number. The Serial
Number must be registered with only one CRM Number.

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ElastographyYes
ConnectivityDICOM, USB, Ethernet, HDMI
Doppler ModesColor Doppler, Power Doppler
Probesconvex, linear, phased array
Imaging ModesM-Mode, Color Doppler, Power Doppler
PortabilityCart-based
ApplicationsVascular, Cardiac, OB/GYN, Musculoskeletal
WeightApprox. 80 kg
Touch Panel10.1-in touchscreen

Related product manuals