GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-137
Section 8-25
Replacement of Battery
8-25-1 Manpower
1 person, 1 hour
NOTE: LP9/P7 BUM 5504583 provides the information, "Refreshing the battery" in Chapter 3 Preparing to
System for Use. The recommended procedure can be executed by user for maintaing battery life.
8-25-2 Tools
Standard Phillips Screwdriver.
8-25-3 Removal Procedure
1) Unscrew 12 screws to remove the IO BOX Cover.
Figure 8-160 Screw points of IO BOX cover