GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
8-122 Section 8-20 - Replacement of Rear Handle
Section 8-20
Replacement of Rear Handle
8-20-1 Manpower
1 Person, 10 minutes.
8-20-2 Tools
Standard Phillips Screwdriver and 6mm Hex Wrench.
8-20-3 Removal Procedure
1) Remove 1 screw and remove the Link Arm Top Cover as following figure.
Figure 8-143 Removal of Link Arm Top Cover.
2) Remove 2 Hex screws. Refer to following figure.
Figure 8-144 Hex screw points of Rear Handle
8-20-4 Installation Procedure
Parts to be installed in reverse order of removal.
NOTE: Hex screws should be tightened.
8-20-5 Functional Check
Visual Check only.
CAUTION
!! CAUTION:
Rear Handle will not stay in position without screws. Make sure not to drop the part when
removing/installing Rear Handle

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ElastographyYes
ConnectivityDICOM, USB, Ethernet, HDMI
Doppler ModesColor Doppler, Power Doppler
Probesconvex, linear, phased array
Imaging ModesM-Mode, Color Doppler, Power Doppler
PortabilityCart-based
ApplicationsVascular, Cardiac, OB/GYN, Musculoskeletal
WeightApprox. 80 kg
Touch Panel10.1-in touchscreen

Related product manuals