Page #461 background imageLoading...
Page #461 background image
GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-135
Section 8-24
Replacement of the HDD/SSD
8-24-1 Manpower
1 Person, 5 minutes.
8-24-2 Tools
Standard Phillips Screwdriver.
8-24-3 Removal Procedure
1) Open the front cover. (refer to section 8-6)
2) Remove 4 screws around HDD/SSD Bracket Refer to following figure.
Figure 8-157 Screw points around HDD/SSD Bracket.
3) Carefully pull down HDD/SSD assy. Refer to following figure.
NOTE: Be careful not to damage HDD/SSD connector
Figure 8-158 Removal of HDD/SSD assy and position of HDD/SSD connector

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ElastographyYes
ConnectivityDICOM, USB, Ethernet, HDMI
Doppler ModesColor Doppler, Power Doppler
Probesconvex, linear, phased array
Imaging ModesM-Mode, Color Doppler, Power Doppler
PortabilityCart-based
ApplicationsVascular, Cardiac, OB/GYN, Musculoskeletal
WeightApprox. 80 kg
Touch Panel10.1-in touchscreen