GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
8-164 Section 8-37 - Replacement of the Monitor Arm
Section 8-37
Replacement of the Monitor Arm
8-37-1 Manpower
1 person, 15 minutes.
8-37-2 Tools
Standard Phillips Screwdriver, Nipper, 4 mm Hex Wrench, 3 Cable ties (GEPN: 2160955).
8-37-3 Removal Procedure
1) Remove LCD monitor and LCD Arm Plastic Covers.(refer to section 8-11)
2) Remove Harness Cable Assy and Monitor Arm. (refer to section 8-11-4 1)~4))
8-37-4 Installation Procedure
In reverse order of removal.
8-37-5 Functional Check
1) Visually check LCD monitor
2) LCD Arm Test. (Refer to section 4-3-2-2).