IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
3-10 Section 3-5 - Preparing for Set Up
3-5-2 EMI Protection
This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
Ensure that the system is protected from electromagnetic interference (EMI), as follows:
Operate the system at least 15 feet away from equipment that emits strong electromagnetic
radiation.
Operate the system in an area enclosed by walls, floors and ceilings comprised of wood, plaster or
concrete, which help prevent EMI.
Shield the system when operating it in the vicinity of radio broadcast equipment, if necessary.
Do not operate mobile phones or other EMI emitting devices in the ultrasound room.
Verify that all EMI rules listed in the following table are followed:
The LOGIQ P9/P7
ultrasound unit is approved for use in hospitals, clinics and other environmentally
qualified facilities, in terms of the prevention of radio wave interference. Operation of the ultrasound unit
in an inappropriate environment can cause electronic interference to radios and television sets situated
near the medical equipment.
For further details and EMI Prevention/Abatement refer to Section 2-2-3 "EMI Limitations" on page 2-4.

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ElastographyYes
ConnectivityDICOM, USB, Ethernet, HDMI
Doppler ModesColor Doppler, Power Doppler
Probesconvex, linear, phased array
Imaging ModesM-Mode, Color Doppler, Power Doppler
PortabilityCart-based
ApplicationsVascular, Cardiac, OB/GYN, Musculoskeletal
WeightApprox. 80 kg
Touch Panel10.1-in touchscreen

Related product manuals