GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
8-32 Section 8-3 - System Software - Installation/Upgrade Procedure
8-3-7-2 System Software - Installation Procedure (R2.x.x)
1) Prepare the Base System Software DVD.
2) Power ON the scanner and press “DEL”(If AN key is installed) or press left bottom key of trackball
(If there’s no AN Key) to enter BIOS
3) Type “maru” for password.
NOTICE
!! NOTICE:
If there’s no AN key, please use trackball to move menu in BIOS mode and click left bottom key of
trackball to select menu in BIOS and double-click to set or change the value in BIOS. Right bottom key
of trackball is ‘ESC’ in BIOS.
Left bottom key of trackball to
enter BIOS and select menu and set
or change the value in BIOS
Right bottom key of trackball
for ‘ESC’ in BIOS