8-114 Section 8-17 - Replacement around Nest Box
8-17-14 Separating Battery Unit from IO BOX
1) Unscrew 12 screws to remove the MIO Cover. (refer to section 8-17-13)
2) Unscrew 8 screws and 2 Battery connectors. Refer to following figure.
Figure 8-130 Screw points and Connectors on battery assy
3) Carefully lift up the battery unit