GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 4 - Functional Checks 4-28
4-3-7-4 Pulse Doppler Mode
Figure 4-26
1) From Utility -> Imaging -> General, verify check on “Simultaneous”.
2) Press B Mode button.
3) Press PW Mode button. (Active PW Mode button on "User Configurable Mode Rotary".)
4) Verify that system scan B/PW duplex with the activated probe.
5) Revert to original setting in Utility -> Imaging -> General “Simultaneous”.
Figure 4-27 Pulse Doppler Mode
4-3-7-5 B/CF/PW Mode
1) Press B Mode button.
2) Press PW Mode Button.
3) Press CF Mode Button.