If the leakage current measurement tests fail on a new unit and if situation cannot be corrected, submit
a Safety Failure Report to document the system problem. Remove unit from operation.
ELECTRICAL SAFETY
Outlet and Ground Continuity test
Earth leakage current test
Enclosure leakage current test
Table 10-20
Outlet/Ground Continuity
Max Value
Allowed
Value
Measured
OK? Comments
Outlet (correct ground & wiring config.) N/A N/A
System Ground Continuity 0.2 Ohm
Table 10-21
Tester Neutral
Tester Polarity
Switch
Max Value Allowed
Value
Measured
OK? Comments
Enter Name of tested peripheral here:
Closed Normal 0.3mA(USA)/0.5mA(Others)
Closed Reversed 0.3mA(USA)/0.5mA(Others)
Open Normal 1mA
Open Reversed 1mA
Table 10-22
Tester
Neutral
Tester
Polarity
Switch
Protective
Earth
Switch
Max Value
Allowed
Test 1
Wheel
Test 2
LCD
Optional
Test 3
Optional
Test 4
OK? Comments
Enter Name of tested peripheral here:
Closed Normal Closed
0.1mA(USA)/
0.1mA(Others)
Closed Normal Open
0.3mA(USA)/
0.5mA(Others)
Closed Reversed Closed
0.1mA(USA)/
0.1mA(Others)
Closed Reversed Open
0.3mA(USA)/
0.5mA(Others)
Open Normal Closed
0.5mA(USA)/
0.5mA(Others)
Open Reversed Closed
0.5mA(USA)/
0.5mA(Others)

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ElastographyYes
ConnectivityDICOM, USB, Ethernet, HDMI
Doppler ModesColor Doppler, Power Doppler
Probesconvex, linear, phased array
Imaging ModesM-Mode, Color Doppler, Power Doppler
PortabilityCart-based
ApplicationsVascular, Cardiac, OB/GYN, Musculoskeletal
WeightApprox. 80 kg
Touch Panel10.1-in touchscreen

Related product manuals