GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-103
8-17-5 Removal Procedure - Nest Box
NOTE: Before working with Nest box, be sure to remove probes.
1) Remove front cover. (refer to section 8-6)
2) Disconnect 1 fan connector of bottom fan.
3) Unlock cable clamp, then pull out connector and cable of bottom fan. Refer to following figures.
Figure 8-109 Fan Connector
4) Remove 2 screws on front side.
Figure 8-110 Front side screw points
5) Remove the rear cover. (refer to section 8-9)
6) Remove Power cord fix bracket and power cord. (refer to section 8-17)
7) Remove cable cover bracket and disconnect cables. (refer to section 8-17)