GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-13
7-4-3-5-3 Log message of System error
Log message format
“Error ; EAErrorHandler(****); Error from” + ErrorModuleName + ErrorMessage
7-4-3-5-4 Types
7-4-3-5-4-1
Example1 Hardware Errors:
Please see the below tables for the details.
Exception Types:
Sub Type Codes:
GFI Hardware Exception
GFI Setup Exception
GFI Buffer Exception
GFI Processing Exception
Code Error Name Description
31 eSwHardwareException_ErrorHandler_GFIHardwareError Hardware Error (MCB/ MDC/MPI..)
167 eSwProcessingException_TsLevelOk_Timeout HV (High Voltage) is not available from MDC.
Hardware error
“Error ; EAErrorHandler(3276); Error from HouseKeeping : GFI Hardware Exception:
SubType 31: -GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_HVLV). Severity :
FatalHW”
Error Module Name
Error descrition
Exception Type
Sub Type Code

Table of Contents

Other manuals for GE LOGIQ P9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P9 and is the answer not in the manual?

GE LOGIQ P9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ElastographyYes
ConnectivityDICOM, USB, Ethernet, HDMI
Doppler ModesColor Doppler, Power Doppler
Probesconvex, linear, phased array
Imaging ModesM-Mode, Color Doppler, Power Doppler
PortabilityCart-based
ApplicationsVascular, Cardiac, OB/GYN, Musculoskeletal
WeightApprox. 80 kg
Touch Panel10.1-in touchscreen

Related product manuals