How to fix 'No connection to Defibrillator Unit' error on GS Medical Equipment?
- EEugene GonzalezAug 26, 2025
This error message can appear if the defibrillator/pacer is outside the radio range. Ensure the defibrillator module is switched on and reduce the distance between the module or connect the modules mechanically. Also, check for any foreign bodies in front of the infrared interface of the individual modules and remove them. If necessary, clean the infrared interfaces.