Keysight E4980A/AL Precision LCR Meter 179
Save/Recall
Saving Measurement Results into USB Memory
Data B
Outputs the measurement data for the secondary parameter.
The Data B field uses the following two different fixed-length ASCII formats:
Status
Represents the measurement result status by using one of the following status
values:
The Status field uses the following two-character fixed-length ASCII format:
SN (S:+/-, N:0 to 4)
Bin No. (IN/OUT)
Represents the bin sorting results as well as IN/OUT evaluation results in list
sweep measurement, as shown below:
When the “effective digits”
setting is on:
(:FORM:ASC:LONG ON)
SN.NNNNNNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
When the “effective digits”
setting is off:
:FORM:ASC:LONG OFF
SN.NNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
When the “effective digits”
setting is on:
:FORM:ASC:LONG ON
SN.NNNNNNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
0 Measurement successfully completed.
+1 Overload.
+3 A signal is detected exceeding the allowable limit of the signal source.
+4 The automatic level control (ALC) feature does not work.
If the value is 1, the measurement data is 9.9E37; if the value is 0, 3, or 4,
the actual measurement data is output.
The data buffer memory contains a measurement result with no data (as
represented by the status value of -1), but it is excluded when the data is
saved into USB memory.
0OUT_OF_BINS
+1 to +9 BIN 1 through BIN 9