- 476 Keysight E4980A/AL User’s Guide
Handler Interface
Setting Up the Handler Interface Board
-
Figure E-9 Flow chart for determining the jumper and bit switch settings
Step 6. Set up the jumpers as instructed in
Table E-11.
Table E-11 Jumper and bit switch settings for test result output signals
Setting # JP1 Bit 5 of
S1
Circuit common Pull-up resistor
mounting
Pull-up voltage
Setting 1 1 On Instrument’s circuit
common
Required +5 V, internal
Setting 2 2 On Instrument’s circuit
common
Required +12 V, internal
Setting 3 3 Off COM1 Required +5 V through +24 V,
external
(EXT.DCV1)
Setting 4 - Off COM1 Not required -
㪼㪋㪐㪏㪇㪸㫌㪼㪈㪈㪋㪉
㪪㪫㪘㪩㪫
㪫㪿㪼㩷㫇㫌㫃㫃㪄㫌㫇㩷㫍㫆㫃㫋㪸㪾㪼㩷㪽㫆㫉㩷㫋㪿㪼㩷㫆㫌㫋㫇㫌㫋㩷㫊㫀㪾㫅㪸㫃㫊㩷㪑
㩷㩷㩷㩷㩷㩷㩷㩿㪈㪀㩷㩷㪬㫊㪼㩷㫋㪿㪼㩷㫀㫅㫋㪼㫉㫅㪸㫃㩷㫍㫆㫃㫋㪸㪾㪼
㩷㩷㩷㩷㩷㩷㩷㩿㪉㪀㩷㩷㪬㫊㪼㩷㫋㪿㪼㩷㪼㫏㫋㪼㫉㫅㪸㫃㩷㫍㫆㫃㫋㪸㪾㪼
㪧㫌㫃㫃㪄㫌㫇㩷㪭㫆㫃㫋㪸㪾㪼㩷㪑
㩷㩷㩷㩷㩷㩷㩷㩿㪈㪀㩷㩷㪬㫊㪼㩷㪂㪌㪭
㩷㩷㩷㩷㩷㩷㩷㩿㪉㪀㩷㩷㪬㫊㪼㩷㪂㪈㪉㪭
㩷㩷㪧㫌㫃㫃㪄㫌㫇㩷㪩㪼㫊㫀㫊㫋㫆㫉㫊㩷㪑
㩿㪈㪀㩷㩷㪤㫆㫌㫅㫋㩷㫆㫅㩷㫋㪿㪼㩷㪿㪸㫅㪻㫃㪼㫉
㩷㩷㩷㩷㩷㩷㩷㩷㩷㩷㪠㪆㪝㩷㪹㫆㪸㫉㪻
㩿㪉㪀㩷㩷㪤㫆㫌㫅㫋㩷㫆㫅㩷㫋㪿㪼㩷㪿㪸㫅㪻㫃㪼㫉
㪪㪼㫋㫋㫀㫅㪾㩷㪈 㪪㪼㫋㫋㫀㫅㪾㩷㪉 㪪㪼㫋㫋㫀㫅㪾㩷㪊 㪪㪼㫋㫋㫀㫅㪾㩷㪋
䋨㪈䋩 䋨㪉䋩
䋨㪈䋩 䋨㪉䋩䋨㪈䋩䋨㪉䋩