Gocator Line Profile Sensors: User Manual
Safety and Maintenance • 17
For more information, see Precautions and Responsibilities below.
Precautions and Responsibilities
Precautions specified in IEC 60825-1 and FDA CFR Title 21 Part 1040 are as follows:
Requirement Class 2 Class 2M Class 3R Class 3B
Remote interlock Not required Not required Not required Required*
Key control Not required Not required Not required Required – cannot
remove key when in
use*
Power-on delays Not required Not required Not required Required*
Beam attenuator Not required Not required Not required Required*
Emission
indicator
Not required Not required Not required Required*
Warning signs Not required Not required Not required Required*
Beam path Not required Not required Terminate beam at
useful length
Terminate beam at
useful length
Specular
reflection
Not required Not required Prevent unintentional
reflections
Prevent unintentional
reflections
Eye protection Not required Not required Not required Required under
special conditions
Laser safety
officer
Not required Not required Not required Required
Training Not required Not required Required for operator
and maintenance
personnel
Required for operator
and maintenance
personnel
*LMI Class 3B laser components do not incorporate these laser safety items. These items must be added and completed by customers
in their system design. For more information, see Class 3B Responsibilities below.
Class 3B Responsibilities
LMI Technologies has filed reports with the FDA to assist customers in achieving certification of laser
products. These reports can be referenced by an accession number, provided upon request. Detailed
descriptions of the safety items that must be added to the system design are listed below.
Remote Interlock
A remote interlock connection must be present in Class 3B laser systems. This permits remote switches
to be attached in serial with the keylock switch on the controls. The deactivation of any remote switches
must prevent power from being supplied to any lasers.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LMI Technologies Gocator 2500 Series and is the answer not in the manual?

LMI Technologies Gocator 2500 Series Specifications

General IconGeneral
BrandLMI Technologies
ModelGocator 2500 Series
CategoryAccessories
LanguageEnglish

Related product manuals