Gocator Line Profile Sensors: User Manual
Protocols • 599
Field Type Offset Description
length 32u 0 Reply size including this field, in bytes.
id 16u 4 Reply identifier (0x1013).
status 32s 6
Reply status. For a list of status codes, see Commands on page
571.
length 32u 10 Data length.
data[length] byte 14 Data content.
Reply
Restore
The Restore command uploads a backup file to the connected sensor and then restores all sensor files
from the backup.
The sensor must be reset or power-cycled before the restore operation can be completed.
Field Type Offset Description
length 32u 0 Command size including this field, in bytes.
id 16u 4 Command identifier (0x1014).
length 32u 6 Data length.
data[length] byte 10 Data content.
Command
Field Type Offset Description
length 32u 0 Reply size including this field, in bytes.
id 16u 4 Reply identifier (0x1014).
status 32s 6
Reply status. For a list of status codes, see Commands on page
571.
Reply
Restore Factory
The Restore Factory command restores the connected sensor to factory default settings.
The command erases the non-volatile memory of the main device.
This command has no effect on connected Buddy sensors.
Note that the sensor must be reset or power-cycled before the factory restore operation can be
completed.
Field Type Offset Description
length 32u 0 Command size including this field, in bytes.
id 16u 4 Command identifier (0x4301).
Command