Gocator Line Profile Sensors: User Manual
Protocols • 660
OK,M00,00,V151290,M01,01,V18520
Standard formatted measurement data with a non-existent measurement of ID 2:
Value,2
ERROR,Specified measurement ID not found. Please verify your input
Custom formatted data string (%time, %value[0]):
Value
OK, 1420266101, 151290
Decision
The Decision command retrieves measurement decisions.
Message Format
Command Decision,measurement ID,measurement ID...
Reply If no arguments are specified, the custom format data string is used.
OK, <custom data string> ERROR, <Error Message>
If arguments are specified,
OK, <data string in standard format, except that the values are not sent> ERROR, <Error
Message>
Formats
Examples:
Standard data string for measurements ID 0 and 1:
Decision,0,1
OK,M00,00,D0,M01,01,D0
Standard formatted measurement data with a non-existent measurement of ID 2:
Decision,2
ERROR,Specified measurement ID not found. Please verify your input
Custom formatted data string (%time, %decision[0]):
Decision
OK,1420266101, 0
Health Channel
The following sections list the actions available on the health channel.