Gocator Line Profile Sensors: User Manual
Safety and Maintenance • 18
Key Control
A key operated master control to the lasers is required that prevents any power from being supplied to
the lasers while in the OFF position. The key can be removed in the OFF position but the switch must not
allow the key to be removed from the lock while in the ON position.
Power-On Delays
A delay circuit is required that illuminates warning indicators for a short period of time before supplying
power to the lasers.
Beam Attenuators
A permanently attached method of preventing human access to laser radiation other than switches,
power connectors or key control must be employed.
Emission Indicator
It is required that the controls that operate the sensors incorporate a visible or audible indicator when
power is applied and the lasers are operating. If the distance between the sensor and controls is more
than 2 meters, or mounting of sensors intervenes with observation of these indicators, then a second
power-on indicator should be mounted at some readily-observable position. When mounting the
warning indicators, it is important not to mount them in a location that would require human exposure
to the laser emissions. User must ensure that the emission indicator, if supplied by OEM, is visible when
viewed through protective eyewear.
Warning Signs
Laser warning signs must be located in the vicinity of the sensor such that they will be readily observed.
Examples of laser warning signs are as follows:
FDA warning sign example IEC warning sign example
Nominal Ocular Hazard Distance (NOHD)
Nominal Ocular Hazard Distance (NOHD)is the distance from the source at which the intensity or the
energy per surface unit becomes lower than the Maximum Permissible Exposure (MPE) on the cornea
and on the skin.
The laser beam is considered dangerous if the operator is closer to the source than the NOHD.
The following tables provide the NOHDvalues for each Gocator model and laser class, assuming
continuous operation of the laser. As a configurable device, Gocator lets you set the laser exposure (laser

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LMI Technologies Gocator 2500 Series and is the answer not in the manual?

LMI Technologies Gocator 2500 Series Specifications

General IconGeneral
BrandLMI Technologies
ModelGocator 2500 Series
CategoryAccessories
LanguageEnglish

Related product manuals