4 Patient Alarms and INOPs
78
* OtherMon Alarm
** OtherMon Alarm
*** OtherMon Alarm
another assigned
monitor
Another monitor used for the patient is in an
alarm condition. Check the detailed alarm
information at the Information Center.
yellow or red alarm lamp and alarm
tone
* Pacer Not Capt ECG/Arrhythmia
(paced patients
only)
No beat detected for > 1.75*(the averaged R-R
interval) with detected pace pulse(s). (For paced
patients only).
numeric flashes, yellow alarm lamp,
short yellow audible alarm
* Pacer Not Pacing ECG/Arrhythmia
(paced patients
only)
No beat and pace pulse detected for >
1.75*(the averaged R-R interval). (For paced
patients only).
numeric flashes, yellow alarm lamp,
short yellow audible alarm
* Pair PVCs ECG/Arrhythmia Two ventricular beats between two non-
ventricular beats.
numeric flashes, yellow alarm lamp,
short yellow audible alarm
* Pause ECG/Arrhythmia No beat detected for a period > the pause
threshold.
numeric flashes, yellow alarm lamp,
short yellow audible alarm
*** <Press Label>
Disconnect
PRESS The pressure is non-pulsatile and the mean
pressure is continuously less than 10 mmHg
(1.3 kPa). This alarm occurs only with arterial
pres su re s (P, AB P, A RT, A o, BA P, FAP, PA P,
UAP, P1, P2, P3, P4).
numeric flashes, red alarm lamp,
alarm tone
*** <Press Label> High PRESS The measured pressure value is above the
extreme high alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes, high limit is
highlighted, red alarm lamp, alarm
tone
** <Press Label> High PRESS The measured pressure value is above the high
alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes, high limit is
highlighted, yellow alarm lamp,
alarm tone
*** <Press Label> Low PRESS The measured pressure value is below the
extreme low alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes and low limit is
highlighted, red alarm lamp, alarm
tone
** <Press Label> Low PRESS The measured pressure value is below the low
alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
** Pulse High PRESS, SpO
2
The pulse rate has exceeded the high alarm
limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** Pulse Low PRESS, SpO
2
The pulse rate has dropped below the low
alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
* PVCs/min High ECG/Arrhythmia PVCs detected within a minute > the alarm
limit
numeric flashes, yellow alarm lamp,
short yellow audible alarm
** QTc High ECG/QT QTc value has exceeded the QTc high limit for
more than 5 minutes
numeric flashes, yellow alarm lamp,
alarm tone
** ΔQTc High ECG/QT ΔQTc value has exceeded the ΔQTc high limit
for more than 5 minutes
numeric flashes, yellow alarm lamp,
alarm tone
Alarm Message From Condition Indication