4 Patient Alarms and INOPs
79
* R-on-T PVCs ECG/Arrhythmia For HR < 100, a PVC with R-R interval < 1/3
the average interval followed by a
compensatory pause of 1.25*(the average R-R
interval), or two such Vs without compensatory
pause occurring within 5 minutes of each other.
(When HR > 100, 1/3 R-R interval is too short
for beat detection)
numeric flashes, yellow alarm lamp,
short yellow audible alarm
** RR High RESP The respiration rate has exceeded the high
alarm limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** RR Low RESP The respiration rate has dropped below the low
alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
* Run PVCs High ECG/Arrhythmia A run of > 2 PVCs. numeric flashes, yellow alarm lamp,
short yellow audible alarm
** <SO₂ Label> High SvO
2
/SO
2
The the measured intravascular oxygen
saturation has exceeded the high limit.
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
** <SO₂ Label> Low SvO
2
/SO
2
The measured intravascular oxygen saturation
has fallen below the low limit.
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
** <SpO₂ Label> High SpO
2
The arterial oxygen saturation has exceeded the
high alarm limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** <SpO₂ Label> Low SpO
2
The arterial oxygen saturation has fallen below
the low alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
**ST Multi <n>,<n> ECG/ST Two contiguous ST leads <n> and <n> have
exceeded elevation or depression limits for
more than 60 seconds. The limit violations for
both leads must be with respect to same limit;
either both above the high limit or both below
the low limit.
numeric flashes, yellow alarm lamp,
alarm tone
**ST Multi
at Information Center
ECG/ST The ST depression or elevation is outside of
the limit in two or more leads. Check on the
monitor
(on monitor) numeric flashes,
yellow alarm lamp, alarm tone
**STE <n>,<n> ECG/ST Two contiguous leads <n> and <n> are above
their respective STE limits
numeric flashes, yellow alarm lamp,
alarm tone
** ST-<n> High ECG/ST The ST elevation in lead <n> is higher than the
limit. Lead is not contiguous with any other
lead.
numeric flashes and high alarm
limit is highlighted, yellow alarm
lamp, alarm tone
** ST-<n> Low ECG/ST The ST depression in lead <n> is lower than
the limit. Lead is not contiguous with any other
lead.
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
* SVT ECG/Arrhythmia A run of supraventricular beats > the SVT run
limit and heart rate > the SVT HR limit.
numeric flashes, yellow alarm lamp,
alarm tone
Alarm Message From Condition Indication