Spectrum Measurements
R&S
®
ESR
412User Manual 1175.7068.02 ─ 12
"Next Min" on page 205
"Next Mode" on page 205
"Search Limits" on page 206
"Peak Excursion" on page 207
Spectrogram: Chapter 5.4.2.3, "Markers in Spectrograms", on page 207
Functions to position markers exclusive in Spectrum mode
Center =Mkr Freq (span > 0).......................................................................................412
Ref Lvl =Mkr Lvl.......................................................................................................... 412
Auto Max Peak/Auto Min Peak................................................................................... 412
Exclude LO..................................................................................................................412
Center =Mkr Freq (span > 0)
Sets the center frequency to the current marker or delta marker frequency. A signal
can thus be set to as center frequency, for example to examine it in detail with a
smaller span.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:CENTer on page 688
Ref Lvl =Mkr Lvl
Sets the reference level to the current marker level.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence on page 862
Auto Max Peak/Auto Min Peak
Adds an automatic peak search action for marker 1 at the end of each particular
sweep. This function may be used during adjustments of a device under test to keep
track of the current peak marker position and level.
The current marker search limit settings (Left Limit, Right Limit, Threshold softkeys) are
taken into account.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:AUTO on page 863
CALCulate<n>:MARKer<m>:MINimum:AUTO on page 863
Exclude LO
Switches the frequency range limit for the marker search functions on or off.
"ON"
The minimum frequency included in the peak search range is ≥ 5 ×
resolution bandwidth (RBW).
Due to the interference by the first local oscillator to the first inter-
mediate frequency at the input mixer, the LO is represented as a sig-
nal at 0 Hz. To avoid the peak marker jumping to the LO signal at 0
Hz, this frequency is excluded from the peak search.
"OFF"
No restriction to the search range. The frequency 0 Hz is included in
the marker search functions.
Remote command:
CALCulate<n>:MARKer<m>:LOEXclude on page 863