Remote Commands
R&S
®
FSW
1007User Manual 1173.9411.02 ─ 43
Parameters:
<PeakPower> ON | OFF | 0 | 1
Turns peak power evaluation on and off.
*RST: 1
<RMSPower> ON | OFF | 0 | 1
Turns RMS power evaluation on and off.
*RST: 0
<AVGPower> ON | OFF | 0 | 1
Turns average power evaluation on and off.
*RST: 0
<TriggerSource> EXTernal | EXT2 | EXT3 | IMMediate | IFPower | RFPower |
VIDeo
Selects a trigger source.
For more information see Configuring Triggered and Gated Mea-
surements.
<TriggerSlope> POSitive | NEGative
Selects the trigger slop.
<TriggerOffset> Defines the trigger delay.
Range: negative measurement time to 30 s
*RST: 0
Default unit: s
<GateLength> Defines the gate length for gated measurements.
Setting 0 seconds turns gated measurements off.
To perform gated measurements, the trigger source must be dif-
ferent from IMMediate.
Range: 31.25 ns to 30 s
*RST: 0 s
Default unit: s
[SENSe:]LIST:POWer:STATe <State>
This command turns the List Evaluation off.
Parameters:
<State> OFF | 0
*RST: 0
14.5.15.2 Example: Performing List Evaluation
The following example shows a list evaluation with the following configuration.
Configuring and Performing Measurements