Common Analysis and Display Functions
R&S
®
FSW
531User Manual 1173.9411.02 ─ 43
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT on page 1166
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt on page 1166
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT on page 1166
Search Minimum
Sets the selected marker/delta marker to the minimum of the trace. If no marker is
active, marker 1 is activated.
For spectrogram displays, define which frame the minimum is to be searched in.
Remote command:
CALCulate<n>:MARKer<m>:MINimum[:PEAK] on page 1165
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK] on page 1167
Search Next Minimum
Sets the selected marker/delta marker to the next (higher) minimum of the selected
trace. If no marker is active, marker 1 is activated.
For spectrogram displays, define which frame the next minimum is to be searched in.
Remote command:
CALCulate<n>:MARKer<m>:MINimum:NEXT on page 1165
CALCulate<n>:MARKer<m>:MINimum:LEFT on page 1164
CALCulate<n>:MARKer<m>:MINimum:RIGHt on page 1165
CALCulate<n>:DELTamarker<m>:MINimum:NEXT on page 1167
CALCulate<n>:DELTamarker<m>:MINimum:LEFT on page 1167
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt on page 1168
Center Frequency = Marker Frequency
Sets the center frequency to the selected marker or delta marker frequency. A peak
can thus be set as center frequency, for example to analyze it in detail with a smaller
span.
This function is not available for zero span measurements.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:CENTer on page 1024
Reference Level = Marker Level
Sets the reference level to the selected marker level.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence on page 1039
9.3.4 Marker (Measurement) Functions
Access: "Overview" > "Analysis" > "Marker Functions"
Marker Usage