Remote Commands
R&S
®
FSW
1311User Manual 1173.9411.02 ─ 43
14.10.11 Remote Commands for Synchronizing Parameters
The commands for manual operation are described in Chapter 12.8, "Synchronizing
Measurement Channel Configuration", on page 719
● Predefined Parameter Coupling..........................................................................1311
● User-Defined Parameter Coupling......................................................................1317
14.10.11.1 Predefined Parameter Coupling
INSTrument:COUPle:ABIMpedance............................................................................... 1311
INSTrument:COUPle:ACDC.......................................................................................... 1312
INSTrument:COUPle:ATTen.......................................................................................... 1312
INSTrument:COUPle:AUNit...........................................................................................1312
INSTrument:COUPle:BANDwidth...................................................................................1313
INSTrument:COUPle:BWIDth........................................................................................ 1313
INSTrument:COUPle:CENTer........................................................................................1313
INSTrument:COUPle:DEMod........................................................................................ 1313
INSTrument:COUPle:GAIN........................................................................................... 1314
INSTrument:COUPle:IMPedance...................................................................................1314
INSTrument:COUPle:LIMit............................................................................................ 1314
INSTrument:COUPle:LLINes.........................................................................................1315
INSTrument:COUPle:MARKer.......................................................................................1315
INSTrument:COUPle:PRESel........................................................................................1315
INSTrument:COUPle:RLEVel........................................................................................ 1316
INSTrument:COUPle:SPAN...........................................................................................1316
INSTrument:COUPle:VBW............................................................................................1317
INSTrument:COUPle:ABIMpedance <State>
This command turns synchronization of the amplitude baseband impedance configura-
tion between measurement channels on and off.
Parameters:
<State> ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST: NONE
Example:
INST:COUP:ABIM ALL
Manual operation: See "Synchronizing parameters across all measurement chan-
nels" on page 720