Data Management
R&S
®
FSW
622User Manual 1173.9411.02 ─ 43
11 Data Management
The R&S FSW allows you to store and load instrument settings, as well as import and
export measurement data for analysis later. Finally, you can store or print the measure-
ment results displayed on the screen.
General storage and import/export functions are available via the toolbar. Some special
storage functions are (also) available via softkeys or dialog boxes in the corresponding
menus, for example trace data export.
See Chapter 9.6, "Importing and Exporting Measurement Results for Evaluation",
on page 608 for RF measurements in the Spectrum mode, or the description of the
specific applications.
Restoring the Default Instrument Configuration (Preset)...................................... 622
Protecting Data Using the Secure User Mode...................................................... 623
Storing and Recalling Instrument Settings and Measurement Data..................... 625
Import/Export Functions........................................................................................636
Creating Screenshots of Current Measurement Results and Settings................. 640
11.1 Restoring the Default Instrument Configuration (Pre-
set)
When delivered, the R&S FSW has a default configuration. You can restore this
defined initial state at any time as a known starting point for measurements. This is
often recommendable as a first step in troubleshooting when unusual measurement
results arise.
Factory default configuration
The factory default configuration is selected such that the RF input is always protected
against overload, provided that the applied signal levels are in the allowed range for
the instrument.
Alternatively to the factory default settings, you can define user-specific recall settings
to be restored after a preset or reboot, see "To recall settings automatically after preset
or reboot" on page 636.
To restore the default instrument configuration for all channels at once
Press the [PRESET] key.
After you use the [PRESET] function, the history of previous actions is deleted, i.e. any
actions performed previously cannot be undone or redone using the [UNDO/REDO]
keys.
Remote command:
*RST or SYSTem:PRESet
Restoring the Default Instrument Configuration (Preset)