Import
Opens a file selection dialog box and loads the transducer factor from the selected file
in .CSV format.
Note that a valid import file must contain a minimum of required information for the
R&S FSW. For details on the file format see Chapter 12.3.3, "Reference: Transducer
Factor File Format", on page 683.
Remote command:
MMEMory:LOAD<n>:TFACtor on page 1271
File Explorer ← Import
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
Export
Opens a file selection dialog box and stores the currently displayed transducer factor to
the defined file in .CSV format.
For details on the file format see Chapter 12.3.3, "Reference: Transducer Factor File
Format", on page 683.
The transducer factor can be imported again later by the R&S FSW for use in other
measurements.
Remote command:
MMEMory:STORe<n>:TFACtor on page 1271
File Explorer ← Export
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
12.3.3 Reference: Transducer Factor File Format
Transducer factor data can be exported to a file in ASCII (CSV) format for further eval-
uation in other applications. Transducer factors stored in the specified ASCII (CSV) for-
mat can also be imported to the R&S FSW for other measurements.
For more information about transducer factors, see "Import" on page 683.
This reference describes in detail the format of the export/import files for transducer
factors. Note that the bold data is mandatory, all other data is optional.
Different language versions of evaluation programs may require a different handling of
the decimal point. Thus, you can define the decimal separator to be used (see "Deci-
mal Separator" on page 612).