Optimizing Measurements
R&S
®
FSW
620User Manual 1173.9411.02 ─ 43
10 Optimizing Measurements
10.1 Minimizing the Measurement Duration
If you want to minimize the measurement duration, try the following methods to opti-
mize the measurement:
●
Reduce the span of the measurement to the relevant parts of the signal only.
●
Increase the RBW to minimize the measurement time; however, consider the
requirements of the standard if you need to measure according to standard!
●
Take advantage of the speed optimization mode in the "Sweep" settings if you do
not require the larger dynamic range (see "Optimization" on page 470).
●
Reduce the sweep time and thus the amount of data to be captured and calcula-
ted; however, consider the requirements regarding the standard deviation.
●
To determine average (or peak) values, use an RMS detector (or peak detector)
with a higher sweep time instead of trace averaging to obtain better average
power results in less time. Furthermore, enforce the use of the sweep type "FFT"
(as opposed to "Auto"; in Sweep mode, the averaging effect of the RMS detector
may be less efficient).
●
When performing multiple measurements, use multiple channels to switch
between measurements rather than changing the settings within one channel
repeatedly.
Take advantage of the Sequencer function to switch between multiple measure-
ments automatically or continuously (see also Chapter 6.4, "Running a Sequence
of Measurements", on page 126).
Additional information
An application note discussing measurement speed optimization is available from the
Rohde & Schwarz website:
1EF90: Speeding up Spectrum Analyzer Measurements
10.2 Improving Averaging Results
Instead of trace averaging, use an RMS detector with a higher sweep time to obtain
better average power results in less time. Furthermore, enforce the use of the sweep
type "FFT" (as opposed to "Auto"; in Sweep mode, the averaging effect of the RMS
detector may be less efficient).
In FFT mode, FFTs are calculated per bin and combined using the RMS detector. For
trace averaging, on the other hand, the local oscillator must be switched repeatedly for
each trace, which takes additional time.
Improving Averaging Results