EasyManuals Logo

R&S FSW Series User Manual

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #470 background imageLoading...
Page #470 background image
Common Measurement Settings
R&S
®
FSW
470User Manual 1173.9411.02 ─ 43
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average", "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
For more information, see Chapter 8.5.1.8, "How Much Data is Measured: Sweep
Points and Sweep Count", on page 464.
For spectrogram displays, the sweep count determines how many sweeps are com-
bined in one frame in the spectrogram; that is: how many sweeps the R&S FSW per-
forms to plot one trace in the spectrogram result display. For more details, see "Time
Frames" on page 590.
Remote command:
[SENSe:]SWEep:COUNt on page 1034
[SENSe:]AVERage<n>:COUNt on page 1129
Sweep Points
Defines the number of measured values to be collected during one sweep.
For details see Chapter 8.5.1.8, "How Much Data is Measured: Sweep Points and
Sweep Count", on page 464.
All values from 101 to 100001 can be set. The default value is 1001 sweep points.
For EMI measurements, 200001 sweep points are available.
Remote command:
[SENSe:]SWEep[:WINDow<n>]:POINts on page 1036
Optimization
In FFT mode, several FFT analysis steps are required to cover the entire measurement
span. The span which is covered by one FFT analysis step is called subspan. The sub-
span cannot be defined directly, but it can be optimized according to measurement
requirements.
Note: To determine the number of required subspans for the measurement, see "FFT
Subspans" on page 472.
Bandwidth, Filter and Sweep Configuration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals