Common Measurement Settings
R&S
®
FSW
469User Manual 1173.9411.02 ─ 43
For more information see Chapter 8.5.1.3, "Coupling VBW and RBW", on page 461.
"Sine[1/1]"
"Video Bandwidth" = "Resolution Bandwidth"
This is the default setting for the coupling ratio RBW/VBW and is rec-
ommended if sinusoidal signals are to be measured.
"Pulse[0.1]"
"Video Bandwidth" = 10 x "Resolution Bandwidth"
or
"Video Bandwidth" = "10 MHz" (= max. VBW)
Recommended for pulse signals
"Noise[10]"
"Video Bandwidth" = "Resolution Bandwidth/10"
Recommended for noise measurements
"Manual"
The coupling ratio is defined manually.
The RBW/VBW ratio can be set in the range of 0.001 to 1000.
Remote command:
[SENSe:]BANDwidth:VIDeo:AUTO on page 1033
[SENSe:]BANDwidth:VIDeo:RATio on page 1033
Filter Type
Defines the filter type.
The following filter types are available:
Normal (3dB)
Channel
RRC
5-Pole (not available for sweep type "FFT")
CISPR (6 dB) - requires EMI (R&S FSW-K54) option
MIL Std (6 dB) - requires EMI (R&S FSW-K54) option
For more information see Chapter 8.5.1.6, "Which Data May Pass: Filter Types",
on page 463.
Note: The EMI-specific filter types are available if the EMI (R&S FSW-K54) measure-
ment option is installed, even if EMI measurement is not active. For details see Chap-
ter 7.13.3.1, "Resolution Bandwidth and Filter Types", on page 334.
The RBW filter configured in the bandwidth settings is identical to the filter configured
in the EMI configuration.
Remote command:
[SENSe:]BANDwidth[:RESolution]:TYPE on page 1032
Default Coupling
Sets all coupled functions to the default state ("Auto"). In addition, the ratio "RBW/
VBW" is set to "Sine[1/1]" and the ratio "Span/RBW" to 100.
For more information see Chapter 8.5.1.3, "Coupling VBW and RBW", on page 461.
Remote command:
[SENSe:]BANDwidth[:RESolution]:AUTO on page 1031
[SENSe:]BANDwidth:VIDeo:AUTO on page 1033
[SENSe:]SWEep:TIME:AUTO on page 1037
Bandwidth, Filter and Sweep Configuration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals