Remote Commands
R&S
®
FSW
1300User Manual 1173.9411.02 ─ 43
SYSTem:SHIMmediate:STATe <State>
Determines when the remote commands that change hardware settings on the
R&S FSW are executed.
Regardless of this setting, the firmware automatically sets the hardware when a sweep
is started.
This setting is not changed by the preset function.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Remote commands that cause changes to the hardware are
only executed when the SYSTem:SHIMmediate ONCE com-
mand is executed.
ON | 1
Remote commands are always executed immediately when they
are received by the instrument.
*RST: 1
Example:
SYST:SHIM:STAT ON
...
SYST:SHIM ONCE
Manual operation: See "Set Hardware Immediately" on page 789
14.10.8 Checking the System Configuration
The following commands are required to check the system configuration on the
R&S FSW.
Useful commands for obtaining system information described elsewhere:
DIAGnostic:SERVice:SINFo? on page 1309
Remote commands exclusive to obtaining system information:
DIAGnostic:INFO:CCOunt?...........................................................................................1301
DEVice:INFO:HWBand?............................................................................................... 1301
DIAGnostic:SERVice:BIOSinfo?.................................................................................... 1302
DIAGnostic:SERVice:HWINfo?...................................................................................... 1302
DIAGnostic:SERVice:VERSinfo?................................................................................... 1303
SYSTem:ERRor:CLEar:ALL.......................................................................................... 1303
SYSTem:DFPRint?.......................................................................................................1303
SYSTem:ERRor:CLEar:REMote.................................................................................... 1304
SYSTem:ERRor:EXTended?......................................................................................... 1304
SYSTem:ERRor:LIST?................................................................................................. 1304
SYSTem:ERRor[:NEXT]?..............................................................................................1305
SYSTem:FIRMware:UPDate..........................................................................................1305
SYSTem:FORMat:IDENt...............................................................................................1306
SYSTem:PRESet:COMPatible.......................................................................................1306
SYSTem:SECurity[:STATe]............................................................................................ 1306
Configuring the R&S
FSW

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals