Data Management
R&S
®
FSW
628User Manual 1173.9411.02 ─ 43
QuickSave 1 / ... / QuickSave 10................................................................................ 628
Storage Type (Save only)............................................................................................628
Recall.......................................................................................................................... 628
QuickSave 1 / ... / QuickSave 10
Selects one of the save sets to store the current settings in or to be recalled. At the
time of storage, the "QuickSave 1 / ... / QuickSave 10" placeholder is replaced by a
label indicating the storage date and time and the storage type.
During recall, save sets of type "Instrument" replace the settings of the entire instru-
ment. All other save sets start a new channel with the stored settings.
Note: Saving instrument settings in secure user mode.
Settings that are saved via Quick Save in secure user mode are only available during
the current session. As soon as the power is switched off on the R&S FSW, the data is
cleared (see Chapter 5.1.7, "Protecting Data Using the Secure User Mode",
on page 43).
Storage Type (Save only)
Defines which type of settings are stored in the save set.
"Instrument
with all Chan-
nels"
The instrument settings for all currently active "Channel"s are stored.
"Current Chan-
nel"
Only the instrument settings for the currently selected measurement
"Channel"s are stored.
Recall
Restores the instrument settings as saved in the selected settings file. If the settings
file contains settings for a specific "Channel" only, a new channel with the stored set-
tings is activated, otherwise all "Channel"s and instrument settings are overwritten with
the stored settings.
Note: After you use the "Recall" function, the history of previous actions is deleted, i.e.
any actions performed previously cannot be undone or redone using the [UNDO/
REDO] keys.
Remote command:
MMEMory:LOAD:STATe on page 1238
11.3.2 Configurable Storage and Recall
The more sophisticated storage and recall functions allow you to define which settings
are stored, and where the settings file is stored to. Any settings file can be selected for
recall.
Stored Data Types................................................................................................ 629
Storage Location and Filename............................................................................ 629
Save and Recall Dialog Boxes..............................................................................629
Startup Recall Settings..........................................................................................632
Storing and Recalling Instrument Settings and Measurement Data

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals