Data Management
R&S
®
FSW
632User Manual 1173.9411.02 ─ 43
Note: Secure user mode. In secure user mode, settings that are stored on the instru-
ment are stored to volatile memory, which is restricted to 256 MB. Thus, a "memory
limit reached" error can occur although the hard disk indicates that storage space is
still available.
For details, see Chapter 5.1.7, "Protecting Data Using the Secure User Mode",
on page 43.
Remote command:
MMEMory:STORe<1|2>:STATe on page 1240
MMEMory:STORe<1|2>:STATe:NEXT on page 1240
Recall in New Channel / Recall in Current Channel
Restores the instrument settings as saved in the selected settings file. If the settings
file contains settings for a specific "Channel" only, select "Recall in New Channel" to
activate a new channel with the stored settings. Select "Recall in Current Channel" to
replace the current "Channel" settings.
Note: After you use the "Recall" function, the history of previous actions is deleted, i.e.
any actions performed previously cannot be undone or redone using the [UNDO/
REDO] keys.
Remote command:
MMEMory:LOAD:STATe on page 1238
11.3.2.4 Startup Recall Settings
Access: "Open" icon in the toolbar > "Startup Recall"
Storing and Recalling Instrument Settings and Measurement Data

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals