EasyManua.ls Logo

R&S FSW Series - Reference: GPIB Commands of Emulated HP Models

R&S FSW Series
1450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
1334User Manual 1173.9411.02 ─ 43
SYSTem:REVision[:STRing] <Name>
Sets the response to the REV? query to the defined string (HP emulation only, see
SYSTem:LANGuage on page 1332).
Parameters:
<Name>
Example: Define the system language:
SYST:LANG '8563E'
Query the revision:
REV?
Response:
920528
Set the response to 'NewRevision':
SYST:REV:STR 'NewRevision'
Query the response:
SYST:REV:STR?
Response:
NewRevision
Manual operation:
See "Revision String" on page 792
SYSTem:RSWeep <State>
Controls a repeated sweep of the E1 and MKPK HI HP model commands (for details
on the commands refer to Chapter 14.13.2, "Reference: GPIB Commands of Emulated
HP Models", on page 1334). If the repeated sweep is OFF, the marker is set without
sweeping before.
This command is only available if a HP language is selected using SYSTem:LANGuage
on page 1332
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYSTem:RSW ON
Manual operation: See "Sweep Repeat" on page 791
14.13.2 Reference: GPIB Commands of Emulated HP Models
The R&S FSW analyzer family supports a subset of the GPIB commands of HP mod-
els 8560E, 8561E, 8562E, 8563E, 8564E, 8565E, 8566A, 8566B, 8568A, 8568B and
8594E.
Despite the differences in system architecture and device features, the supported com-
mands have been implemented in a way to ensure a sufficiently high degree of corre-
spondence with the original.
This includes the support of syntax rules for not only newer device families (B and E
models) but for the previous A family as well.
Emulating Other Instruments' Commands

Table of Contents