SYSTem:REVision[:STRing] <Name>
Sets the response to the REV? query to the defined string (HP emulation only, see
SYSTem:LANGuage on page 1332).
Parameters:
<Name>
Example: Define the system language:
SYST:LANG '8563E'
Query the revision:
REV?
Response:
920528
Set the response to 'NewRevision':
SYST:REV:STR 'NewRevision'
Query the response:
SYST:REV:STR?
Response:
NewRevision
Manual operation:
See "Revision String" on page 792
SYSTem:RSWeep <State>
Controls a repeated sweep of the E1 and MKPK HI HP model commands (for details
on the commands refer to Chapter 14.13.2, "Reference: GPIB Commands of Emulated
HP Models", on page 1334). If the repeated sweep is OFF, the marker is set without
sweeping before.
This command is only available if a HP language is selected using SYSTem:LANGuage
on page 1332
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYSTem:RSW ON
Manual operation: See "Sweep Repeat" on page 791
14.13.2 Reference: GPIB Commands of Emulated HP Models
The R&S FSW analyzer family supports a subset of the GPIB commands of HP mod-
els 8560E, 8561E, 8562E, 8563E, 8564E, 8565E, 8566A, 8566B, 8568A, 8568B and
8594E.
Despite the differences in system architecture and device features, the supported com-
mands have been implemented in a way to ensure a sufficiently high degree of corre-
spondence with the original.
This includes the support of syntax rules for not only newer device families (B and E
models) but for the previous A family as well.
Emulating Other Instruments' Commands