General Instrument Setup
R&S
®
FSW
706User Manual 1173.9411.02 ─ 43
If the number of error messages exceeds the capacity of the error buffer, "Message
Buffer Overflow" is displayed. To clear the message buffer use the "Clear All Mes-
sages" button.
The following information is available:
No device-specific error code
Message brief description of the message
Component hardware messages: name of the affected module
software messages: name of the affected software
Date/Time date and time of the occurrence of the message
Remote command:
SYSTem:ERRor:LIST? on page 1304
12.6.4 Firmware Updates
Access: [Setup] > "System Configuration" > "Firmware Update"
During instrument start, the installed hardware is checked against the current firmware
version to ensure the hardware is supported. If not, an error message is displayed
("Wrong Firmware Version") and you are asked to update the firmware. Until the firm-
ware version is updated, self-alignment fails. To see which components are not suppor-
ted, see the System Messages.
The firmware on your R&S FSW may also need to be updated in order to enable addi-
tional new features or if reasons for improvement come up. Ask your sales representa-
System Configuration Settings

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals