Remote Commands
R&S
®
FSW
1295User Manual 1173.9411.02 ─ 43
Suffix Description
CMAP30 Limit and Margin Check – Text
CMAP31 Limit and Margin Check – \"Margin\"
CMAP32 *) Table Overall – Title Text
CMAP33 *) Table Overall – Title Background
CMAP34 *) Table Overall – Text
CMAP35 *) Table Overall – Background
CMAP36 *) Table Value – Title Text
CMAP37 *) Table Value – Title Background
CMAP38 *) Table Value – Text
CMAP39 *) Table Value – Background
CMAP40 Trace 5
CMAP41 Trace 6
*) these settings can only be defined via the theme (DISPlay:THEMe:SELect) and
are thus ignored in the SCPI command
14.10.6 Remote Commands for Language Settings
SYSTem:DISPlay:LANGuage........................................................................................ 1295
SYSTem:DISPlay:LANGuage <Language>
Defines the language of the software-defined interface elements (such as softkeys,
dialog boxes, diagram texts etc.).
Parameters:
<Language> 'EN' | 'ZH_CH' | 'ZH_TW' | 'JA' | 'KO' | 'RU'
'ZH_CH'
Simplified Chinese
'ZH_TW'
Traditional Chinese
*RST: 'EN'
Example:
SYST:DISP:LANG 'JA'
Switches the language of the instrument to Japanese.
14.10.7 Configuring the Network and Remote Control
The following commands are required to configure a network or remote control for the
R&S FSW.
Configuring the R&S
FSW